Back to Search Start Over

Friction control of mechanical seals in a ventricular assist device

Authors :
K. Kanda
H. Sato
T. Miyakoshi
T. Kitano
H. Kanebako
K. Adachi
Source :
Biosurface and Biotribology, Vol 1, Iss 2, Pp 135-143 (2015)
Publication Year :
2015
Publisher :
Wiley, 2015.

Abstract

Low and stable friction is required for mechanical seals in implantable ventricular assist devices. In this study, a specialized test apparatus was designed to test the frictional properties of a mechanical seal in blood in implantable ventricular assist devices. It was shown that a blood-derived protein film forms on the sealing surfaces and causes higher and unstable friction than that in water. Further, it was shown that concave surface features on the substrate initially catch aggregated proteins that are denatured by friction, thus the protein film progresses from concave to flat regions on the substrate. On the basis of this protein film formation mechanism, the creation of a smooth, hydrophilic sealing surface was proposed to control friction and its effectiveness was validated.

Details

Language :
English
ISSN :
24054518
Volume :
1
Issue :
2
Database :
Directory of Open Access Journals
Journal :
Biosurface and Biotribology
Publication Type :
Academic Journal
Accession number :
edsdoj.50948640a09046faa8be8075273541c1
Document Type :
article
Full Text :
https://doi.org/10.1016/j.bsbt.2015.06.004