Interference with pulse oximetry by the Stealth Station™ Image Guidance System
- Source :
- JA Clinical Reports, Vol 3, Iss 1, Pp 1-3 (2017)
- Publication Year :
- 2017
- Publisher :
- SpringerOpen, 2017.
-
Abstract
- Abstract Background A pulse oximeter is one of the most important monitors to save patients undergoing anesthesia and monitored sedation. The authors report a case of orthopedic surgery, in which interference of pulse oximetry occurred when using a Stealth Station™ navigation system (Medtronic Sofamor Danek, Memphis, TN). Applying a black plastic shield (Masimo Ambient Shield: Masimo Corporation, Irvine, CA) completely eliminated the interference. Case presentation A 37-year-old male patient with a giant cell tumor of the left femur was scheduled to undergo curettage of the femur using an intraoperative CT three-dimensional imaging system (O-arm™) and Stealth Station™ navigation system. During the surgery, the SpO2 value, which was maintained between 97 and 99% until the time, disappeared suddenly with abnormal pulse wave. Because a distortion in the SpO2 value was reproduced by repeated movement of cameras on the head of the Stealth Station™ navigation system, we recognized that the interference signal was coming from the navigation system. To eliminate the infrared light, the pulse oximetry probe was covered with a black plastic shield and the interference was completely eliminated. Conclusions The Stealth Station™ navigation system was found to interfere with the SpO2 value, and a black plastic shield was useful for eliminating the interfering signal. Anesthesiologists should understand the risk of interference by the neuro-navigation system and know how to solve the problem.
Details
- Language :
- English
- ISSN :
- 23639024
- Volume :
- 3
- Issue :
- 1
- Database :
- Directory of Open Access Journals
- Journal :
- JA Clinical Reports
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.4f223baf0045d3a39e6e2fea21d540
- Document Type :
- article
- Full Text :
- https://doi.org/10.1186/s40981-017-0076-7