Acquisition of Quasi-Monochromatic Dual-Energy in a Microfocus X-ray Generator and Development of Applied Technology

Authors :
Hiroaki Hasegawa
Masanori Sato
Source :
Diagnostics, Vol 9, Iss 1, p 27 (2019)
Publication Year :
2019
Publisher :
MDPI AG, 2019.

Abstract

In regenerative medicine, evaluation of bone mineral density using a microfocus X-ray generator could eventually be used to determine the degree of bone tissue regeneration. To evaluate bone mineral density against regenerated bone material, two low-energy X-rays are necessary. Herein, the acquisition of quasi-monochromatic, dual-energy soft X-ray and the subsequent medical application were examined using the K-absorption edges of two types of metal filters (i.e., zirconium and tin) in a microfocus X-ray generator. Investigation of the optimal tube voltage and filter thickness to form a quasi-monochromatic energy spectrum with a single filter revealed that a filter thickness of 0.3 mm results in an optimal monochromatization state. When a dual filter was used, the required filter thickness was 0.3 mm for tin and 0.2 mm for zirconium at a tube voltage of 35 kV. For the medical application, we measured quasi-monochromatic, dual-energy X-rays to evaluate the measurement accuracy of bone mineral density. Using aluminum as a simulated bone sample, a relative error of ≤5% was consistent within the aluminum thickness range of 1–3 mm. These data suggest that a bone mineral density indicator of recycled bone material can be easily obtained with the quasi-monochromatic X-ray technique using a microfocus X-ray generator.

Details

Language :
English
ISSN :
20754418
Volume :
9
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Diagnostics
Publication Type :
Academic Journal
Accession number :
edsdoj.4c64b47632da47a0a7fa7af94902e4b8
Document Type :
article
Full Text :
https://doi.org/10.3390/diagnostics9010027