Paper-Based Sensor Chip for Heavy Metal Ion Detection by SWSV
- Source :
- Micromachines, Vol 9, Iss 4, p 150 (2018)
- Publication Year :
- 2018
- Publisher :
- MDPI AG, 2018.
-
Abstract
- Heavy metal ion pollution problems have had a terrible influence on human health and the environment. Therefore, the monitoring of heavy metal ions is of great practical significance. In this paper, an electrochemical three-electrode system was fabricated and integrated on nitrocellulose membrane (NC) by the use of magnetron sputtering technology, which exhibited a uniform arrangement of porous structure without further film modification. This paper-based sensor chip was used for Cu2+ detection by square-wave stripping voltammetry (SWSV). Within the ranges of 5~200 μg·L−1 and 200~1000 μg·L−1, it showed good linearity of 99.58% and 98.87%, respectively. The limit of detection was 2 μg·L−1. On the basis of satisfying the detection requirements (10 μg·L−1), the integrated sensor was small in size and inexpensive in cost. Zn2+, Cd2+, Pb2+ and Bi3+ were also detected by this paper-based sensor chip with good linearity.
Details
- Language :
- English
- ISSN :
- 2072666X
- Volume :
- 9
- Issue :
- 4
- Database :
- Directory of Open Access Journals
- Journal :
- Micromachines
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.46443302257411daf3dde9cec8042f4
- Document Type :
- article
- Full Text :
- https://doi.org/10.3390/mi9040150