Abstract
- Abstract Background The muscle artifacts, caused by prominent muscle contractions, mimicking cardiac arrhythmias, might compromise the ECG signal quality and the implantable loop recorder memory capacity in patients with epilepsy. We developed an epileptic seizures clinical pattern-based implantable loop recorder manual activation algorithm, presenting its real-world efficacy here. Methods One hundred ninety-three patients (18–60 years) with drug-resistant focal epilepsy were consecutively enrolled and underwent a subcutaneous loop recorder implantation. Patients with focal onset-aware seizures and patients with focal impaired awareness seizures /bilateral tonic-clonic seizures without aura were recommended to use the activator once - just after the episode. Patients with focal impaired awareness seizures/bilateral tonic-clonic seizures with aura, the caregivers of patients experiencing status epilepticus, were advised to use the activator twice - during the aura and after the episode/ regaining consciousness. Results Six thousand four hundred ninety-four ECG traces (4826 - auto-triggered events, 1668 - person-activated events) were recorded and analyzed. The rate of true positive events in the person-activated group was statistically higher than in the autoactivation group (72.5% vs.19.4%, p
Details
- Language :
- English
- ISSN :
- 14712261
- Volume :
- 24
- Issue :
- 1
- Database :
- Directory of Open Access Journals
- Journal :
- BMC Cardiovascular Disorders
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.40768f0e5b5481ebfc82f4ed3463eb2
- Document Type :
- article
- Full Text :
- https://doi.org/10.1186/s12872-024-03721-5