Validation of a point-of-care capillary lactate measuring device (Lactate Pro 2)
- Source :
- Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine, Vol 28, Iss 1, Pp 1-7 (2020)
- Publication Year :
- 2020
- Publisher :
- BMC, 2020.
-
Abstract
- Abstract Background The measurement of lactate in emergency medical services has the potential for earlier detection of shock and can be performed with a point-of-care handheld device. Validation of a point-of-care handheld device is required for prehospital implementation. Aim The primary aim was to validate the accuracy of Lactate Pro 2 in healthy volunteers and in haemodynamically compromised intensive care patients. The secondary aim was to evaluate which sample site, fingertip or earlobe, is most accurate compared to arterial lactate. Methods Arterial, venous and capillary blood samples from fingertips and earlobes were collected from intensive care patients and healthy volunteers. Arterial and venous blood lactate samples were analysed on a stationary hospital blood gas analyser (ABL800 Flex) as the reference device and compared to the Lactate Pro 2. We used the Bland-Altman method to calculate the limits of agreement and used mixed effect models to compare instruments and sample sites. A total of 49 intensive care patients with elevated lactate and 11 healthy volunteers with elevated lactate were included. Results There was no significant difference in measured lactate between Lactate Pro 2 and the reference method using arterial blood in either the healthy volunteers or the intensive care patients. Capillary lactate measurement in the fingertip and earlobe of intensive care patients was 47% (95% CI (29 to 68%), p
Details
- Language :
- English
- ISSN :
- 17577241
- Volume :
- 28
- Issue :
- 1
- Database :
- Directory of Open Access Journals
- Journal :
- Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.3cc1e99e70145439a5f03cc28aa0e73
- Document Type :
- article
- Full Text :
- https://doi.org/10.1186/s13049-020-00776-z