Agreement between Vital Signs Measured Using Mat-Type Noncontact Sensors and Those from Conventional Clinical Assessment

Authors :
Daiki Shimotori
Eri Otaka
Kenji Sato
Munetaka Takasugi
Nobuyoshi Yamakawa
Atsuya Shimizu
Hitoshi Kagaya
Izumi Kondo
Source :
Healthcare, Vol 12, Iss 12, p 1193 (2024)
Publication Year :
2024
Publisher :
MDPI AG, 2024.

Abstract

Vital signs are crucial for assessing the condition of a patient and detecting early symptom deterioration. Noncontact sensor technology has been developed to take vital measurements with minimal burden. This study evaluated the accuracy of a mat-type noncontact sensor in measuring respiratory and pulse rates in patients with cardiovascular diseases compared to conventional methods. Forty-eight hospitalized patients were included; a mat-type sensor was used to measure their respiratory and pulse rates during bed rest. Differences between mat-type sensors and conventional methods were assessed using the Bland–Altman analysis. The mean difference in respiratory rate was 1.9 breaths/min (limits of agreement (LOA): −4.5 to 8.3 breaths/min), and proportional bias existed with significance (r = 0.63, p < 0.05). For pulse rate, the mean difference was −2.0 beats/min (LOA: −23.0 to 19.0 beats/min) when compared to blood pressure devices and 0.01 beats/min (LOA: −11.4 to 11.4 beats/min) when compared to 24-h Holter electrocardiography. The proportional bias was significant for both comparisons (r = 0.49, p < 0.05; r = 0.52, p < 0.05). These were considered clinically acceptable because there was no tendency to misjudge abnormal values as normal. The mat-type noncontact sensor demonstrated sufficient accuracy to serve as an alternative to conventional assessments, providing long-term monitoring of vital signs in clinical settings.

Details

Language :
English
ISSN :
12121193 and 22279032
Volume :
12
Issue :
12
Database :
Directory of Open Access Journals
Journal :
Healthcare
Publication Type :
Academic Journal
Accession number :
edsdoj.37fef426171a482db560380f240bccd2
Document Type :
article
Full Text :
https://doi.org/10.3390/healthcare12121193