Back to Search Start Over

How Accurate Are Electronic Monitoring Devices? A Laboratory Study Testing Two Devices to Measure Medication Adherence

Authors :
Fabienne Dobbels
Johan Vanhaecke
Sofie Vandenbroeck
Sabina De Geest
Leentje De Bleser
Source :
Sensors, Vol 10, Iss 3, Pp 1652-1660 (2010)
Publication Year :
2010
Publisher :
MDPI AG, 2010.

Abstract

In a prospective descriptive laboratory study, 25 Helping Handâ„¢ (HH) (10 without and 15 with reminder system) and 50 Medication Event Monitoring Systems (MEMS) (25 with 18-month and 25 with 2-year battery life) were manipulated twice daily following a predefined protocol during 3 consecutive weeks. Accuracy was determined using the fixed manipulation scheme as the reference. Perfect functioning (i.e., total absence of missing registrations and/or overregistrations) was observed in 70% of the HH without, 87% of the HH with reminder, 20% MEMS with 18 months, and 100% with 2-year battery life respectively.

Details

Language :
English
ISSN :
14248220
Volume :
10
Issue :
3
Database :
Directory of Open Access Journals
Journal :
Sensors
Publication Type :
Academic Journal
Accession number :
edsdoj.373b0ffd53cf42bd967afc2e04c72245
Document Type :
article
Full Text :
https://doi.org/10.3390/s100301652