- Background This in vitro investigation aimed to determine the influence of multiple autoclave cycles on the cyclic fatigue resistance of three heat-treated nickel-titanium (NiTi) files: EdgeTaper Platinum (ETP), ProTaper Gold (PTG), and TruNatomy Prime (TN). Materials Sixty NiTi files, twenty of each NiTi file type: ETP 25/.06, PTG 25/.08, and TN 26/.04 were randomly divided into four equal subgroups (n = 5). The files for the control group were left un-autoclaved. Different autoclave sterilization cycles (one, five, and ten) were used for the other three groups. The files were then placed in a metal canal block and rotated according to the manufacturer’s instructions until fracture. The length of the broken segment and the time taken for fracture were measured. The fractured surfaces were subsequently subjected to SEM imaging. The Kruskal-Wallis test was used to analyze the data, followed by Dunn-Bonferroni pairwise comparisons. Statistical significance was set at p
Details
- Language :
- English
- ISSN :
- 21678359
- Volume :
- 11
- Database :
- Directory of Open Access Journals
- Journal :
- PeerJ
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.33b1139fbe6f40e0b68078bc56262e4f
- Document Type :
- article
- Full Text :
- https://doi.org/10.7717/peerj.14656