Optimization of Parameters in 16-slice CT-‌‌scan Protocols for Reduction of the Absorbed Dose

Authors :
Shahrokh Naseri
Mahdi Momen Nezhad
Zahra Hozhabri
Abbas Haghparast
Golestan Karami
Peyman Hejazi
Source :
Iranian Journal of Medical Physics, Vol 11, Iss Issue 2,3, Pp 270-275 (2014)
Publication Year :
2014
Publisher :
Mashhad University of Medical Sciences, 2014.

Abstract

Introduction In computed tomography (CT) technology, an optimal radiation dose can be achieved via changing radiation parameters such as mA, pitch factor, rotation time and tube voltage (kVp) for diagnostic images. Materials and Methods In this study, the brain, abdomen, and thorax scaning was performed using Toshiba 16-slice scannerand standard AAPM and CTDI phantoms. AAPM phantom was used for the measurement of image-related parameters and CTDI phantom was utilized for the calculation of absorbed dose to patients. Imaging parameters including mA (50-400 mA), pitch factor (1 and 1.5) and rotation time (range of 0.5, 0.75, 1, 1.5 and 2 seconds) were considered as independent variables. The brain, abdomen and chest imaging was performed multi-slice and spiral modes. Changes in image quality parameters including contrast resolution (CR) and spatial resolution (SR) in each condition were measured and determined by MATLAB software. Results After normalizing data by plotting the full width at half maximum (FWHM) of point spread function (PSF) in each condition, it was observed that image quality was not noticeably affected by each cases. Therefore, in brain scan, the lowest patient dose was in 150 mA and rotation time of 1.5 seconds. Based on results of scanning of the abdomen and chest, the lowest patient dose was obtained by 100 mA and pitch factors of 1 and 1.5. Conclusion It was found that images with acceptable quality and reliable detection ability could be obtained using smaller doses of radiation, compared to protocols commonly used by operators.

Details

Language :
English
ISSN :
23453672
Volume :
11
Issue :
Issue 2,3
Database :
Directory of Open Access Journals
Journal :
Iranian Journal of Medical Physics
Publication Type :
Academic Journal
Accession number :
edsdoj.279b65c1ab7490b8602f2ea022f7450
Document Type :
article
Full Text :
https://doi.org/10.22038/ijmp.2014.3101