Back to Search
Start Over
Temperature-induced configuration changes in hydrogel-coated coils and their relevance in embolization procedures
- Source :
- CVIR Endovascular, Vol 3, Iss 1, Pp 1-7 (2020)
- Publication Year :
- 2020
- Publisher :
- SpringerOpen, 2020.
-
Abstract
- Abstract Background The present study attempted to demonstrate how the configuration of hydrogel-coated coils is influenced by different temperature exposures. Thirty detachable hydrogel-coated coils were evaluated in an in vitro water immersion test under five different temperature ranges (22.6 °C, 37 °C, 40–50 °C, 50–60 °C, and 60–70 °C). The configuration changes were classified (configuration I, configuration II, and configuration III) according to the curling that occurred during 30 min of immersion. Configuration stability of five Hydrogel-coated coils was also evaluated in a two-step temperature immersion test. Results All hydrogel-coated coils showed some configuration changes during water immersion. However, a logarithmic transformation of the time and temperature data showed a significant (p
Details
- Language :
- English
- ISSN :
- 25208934
- Volume :
- 3
- Issue :
- 1
- Database :
- Directory of Open Access Journals
- Journal :
- CVIR Endovascular
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.25028645c86d4558a48abd7b920245d2
- Document Type :
- article
- Full Text :
- https://doi.org/10.1186/s42155-020-00189-0