Abstract

Abstract Background The present study attempted to demonstrate how the configuration of hydrogel-coated coils is influenced by different temperature exposures. Thirty detachable hydrogel-coated coils were evaluated in an in vitro water immersion test under five different temperature ranges (22.6 °C, 37 °C, 40–50 °C, 50–60 °C, and 60–70 °C). The configuration changes were classified (configuration I, configuration II, and configuration III) according to the curling that occurred during 30 min of immersion. Configuration stability of five Hydrogel-coated coils was also evaluated in a two-step temperature immersion test. Results All hydrogel-coated coils showed some configuration changes during water immersion. However, a logarithmic transformation of the time and temperature data showed a significant (p