Consistency analysis of Keratograph and traditional methods to evaluate tear film function
- Source :
- Guoji Yanke Zazhi, Vol 15, Iss 5, Pp 846-849 (2015)
- Publication Year :
- 2015
- Publisher :
- Press of International Journal of Ophthalmology (IJO PRESS), 2015.
-
Abstract
- AIM: To investigate repeatability and accuracy of a latest Keratograph for evaluating the tear film stability and to compare its measurements with that of traditional examination methods. METHODS: The results of noninvasive tear film break-up time(NI-BUT)including the first tear film break-up time(BUT-f)and the average tear film break-up time(BUT-ave)were measured by Keratograph. The repeatability of the measurements was evaluated by coefficient of variation(CV)and intraclass correlation coefficient(ICC). Wilcoxon Signed-Rank test was used to compare NI-BUT with fluorescein tear film break-up time(FBUT)to confirm the correlation between NI-BUT and FBUT, Schirmer I test values. Bland-Altman analysis was used to evaluate consistency. RESULTS: The study recruited 48 subjects(48 eyes)(mean age 38.7±15.2 years). The CV and ICC of BUT-f were respectively 12.6% and 0.95, those of BUT-ave were 9.8% and 0.96. The value of BUT-f was lower than that of FBUT. The difference had statistical significance(6.16±2.46s vs 7.46±1.92s, PPCONCLUSION: Keratograph can provide NI-BUT data that has a better repeatability and reliability, which has great application prospects in diagnosis and treatment of dry eye and refractive corneal surgery.
Details
- Language :
- English
- ISSN :
- 16725123
- Volume :
- 15
- Issue :
- 5
- Database :
- Directory of Open Access Journals
- Journal :
- Guoji Yanke Zazhi
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.1a6fb10b845e41d8b980db9e00da3d2c
- Document Type :
- article
- Full Text :
- https://doi.org/10.3980/j.issn.1672-5123.2015.5.27