Comparison of Diagon Coag XL and Cobas t 511 analyzers

Authors :
Berrak Güven
Ismail Benice
Murat Can
Source :
International Journal of Medical Biochemistry, Vol 5, Iss 2, Pp 116-119 (2022)
Publication Year :
2022
Publisher :
KARE Publishing, 2022.

Abstract

INTRODUCTION: We compared prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, and Ddimer test results measured using the Diagon Coag XL coagulation analyzer with Cobas t 511 analyzer. METHODS: Imprecision studies were performed for the PT, aPTT, fibrinogen, and D-dimer tests used by the Diagon Coag XL analyzer. For the comparison study, we used the leftover citrated plasma from patient samples after routine analysis with Cobas t 511. All of the results were analyzed using the correlation coefficient and Passing–Bablok regression analysis. RESULTS: Total coefficient of variation obtained for all tests were within the criteria for acceptance. The method comparison study showed a good correlation between the results obtained on Diagon Coag XL and Cobas t 511 analyzers, except for aPTT test. The correlation coefficients obtained were 0.98 for PT, INR, and D-dimer, 0.95 for fibrinogen, and 0.80 for aPTT. DISCUSSION AND CONCLUSION: For PT, aPTT, fibrinogen, and D-dimer tests, Diagon Coag XL analyzer is suitable for monitoring the coagulation system, and it can be used in clinical laboratories. However, the precision values of tests stated by the manufacturer must be verified.

Details

Language :
English
ISSN :
25872362
Volume :
5
Issue :
2
Database :
Directory of Open Access Journals
Journal :
International Journal of Medical Biochemistry
Publication Type :
Academic Journal
Accession number :
edsdoj.15245d1da6f1488eba5ebb10336b90c9
Document Type :
article
Full Text :
https://doi.org/10.14744/ijmb.2022.40316