Evaluation of Xpert point-of-care assays for detection of HIV infection in persons using long-acting cabotegravir for pre-exposure prophylaxis

Authors :
Jessica M. Fogel
Estelle Piwowar-Manning
Amber Moser
Tinia Hill
Shahnaz Ahmed
Vanessa Cummings
Heba H. Mostafa
Zhe Wang
Andrea Jennings
Jorge A. Gallardo-Cartagena
María Inés Figueroa
Marty St Clair
Alex R. Rinehart
Adeola Adeyeye
James F. Rooney
Myron S. Cohen
Beatriz Grinsztejn
Raphael J. Landovitz
Susan H. Eshleman
Source :
Microbiology Spectrum, Vol 12, Iss 8 (2024)
Publication Year :
2024
Publisher :
American Society for Microbiology, 2024.

Abstract

ABSTRACT Detection of HIV infection may be challenging in persons using long-acting cabotegravir (CAB-LA) pre-exposure prophylaxis (PrEP) due to viral suppression and reduced/delayed antibody production. We evaluated two point-of-care tests for detecting HIV infection in persons who received CAB-LA in the HPTN 083 trial. Samples were obtained from 12 participants who received CAB-LA and had delayed detection of HIV infection using HIV rapid tests and an antigen/antibody test (52 plasma samples; 18 dried blood spot [DBS] samples). Plasma samples were tested with the Xpert HIV-1 Viral Load XC test (Xpert VL-XC); DBS samples were tested with the total nucleic acid Xpert HIV-1 Qual XC test (Xpert Qual-XC). Results from these assays were compared to results from three reference, laboratory-based, plasma RNA assays (Aptima HIV-1 Qualitative assay [Aptima Qual]; Aptima HIV-1 Quant DX Assay [Aptima Quant]; cobas HIV-1/HIV-2 Qualitative Test [cobas]). HIV RNA was detected with all four plasma assays for all samples with viral loads (VLs) ≥ 200 copies/mL; the number of samples with VLs < 200 copies/mL with HIV RNA detected was: Xpert VL-XC: 19/26 (73.1%); Aptima Qual: 17/26 (65.4%); Aptima Quant: 17/26 (65.4%); and cobas: 12/21 (57.1%). The Xpert Qual-XC assay was positive for all DBS samples with VLs ≥ 200 copies/mL and 1/10 DBS with VLs < 200 copies/mL. The performance of the Xpert VL-XC assay was comparable to the reference assays for detecting HIV infection in these cases. The Xpert Qual-XC assay was less sensitive than plasma-based HIV RNA assays for detecting HIV in the setting of CAB-LA PrEP.IMPORTANCEHIV RNA assays can detect HIV infections earlier than HIV rapid tests or Ag/Ab tests in persons using CAB-LA PrEP. Earlier HIV diagnosis could allow for earlier treatment initiation and reduced risk of INSTI resistance. POC tests may help detect HIV infection before CAB-LA administration and may be more accessible than laboratory-based assays in some settings. In this study, the POC Xpert VL-XC assay detected HIV RNA in most samples from individuals who received CAB-LA PrEP and had delayed detection of HIV infection with HIV rapid tests and an Ag/Ab test. The performance of this assay was similar to laboratory-based HIV RNA assays in this cohort. The POC Xpert Qual-XC assay detects both HIV RNA and DNA, with a higher viral load cutoff for RNA detection. This assay was negative for most lower viral load samples and did not offer an advantage for HIV screening in persons using CAB-LA PrEP.

Details

Language :
English
ISSN :
21650497
Volume :
12
Issue :
8
Database :
Directory of Open Access Journals
Journal :
Microbiology Spectrum
Publication Type :
Academic Journal
Accession number :
edsdoj.0ec30651f1934244afc8909a571ed9c4
Document Type :
article
Full Text :
https://doi.org/10.1128/spectrum.00307-24