Back to Search Start Over

Development and validation of a tissue-equivalent test environment for detection of malfunctions in active medical implants caused by ionizing radiation

Authors :
Stich Manuel
Blümlein Larissa
Slawig Anne
Schmidl Felix
Schuller Karina
Lösch Richard
Hipp Matthias
Hentschel Sabine
Schaefers Gregor
Ringler Ralf
Source :
Current Directions in Biomedical Engineering, Vol 4, Iss 1, Pp 153-156 (2018)
Publication Year :
2018
Publisher :
De Gruyter, 2018.

Abstract

Many patients in radiotherapy carry active implantable medical devices (AIMDs) such as pacemakers or cardioverter defibrillators (ICDs). The influence of the ionizing radiation can lead to failures in the device function. This study presents a tissue-equivalent test environment to investigate the influence of ionizing radiation on AIMDs. The in-vitro test environment is designed to simulate a human torso. Structures such as the heart, lungs, ribs, spinal column and soft tissue are replicated from tissue-equivalent materials to allow realistic treatment planning and to simulate the effect of ionizing radiation on active implants. CT measurements and Monte-Carlo validations have shown that Polytetrafluorethylen (bone), carrageenan (heart), Styrodur (lung) and Biresin® G27 (soft tissue) fulfill all requirements for suitable tissue surrogates. A plug-in unit integrated in the test environment has been designed specifically to allow the placement AIMDs in the phantom at typical positions for implant placement in humans. The dosimetry validation showed that the test environment is applicable in the full treatment planning process.

Details

Language :
English
ISSN :
23645504
Volume :
4
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Current Directions in Biomedical Engineering
Publication Type :
Academic Journal
Accession number :
edsdoj.0dab70771cb424ea5bc0657b1596c28
Document Type :
article
Full Text :
https://doi.org/10.1515/cdbme-2018-0038