Back to Search
Start Over
W+Rh as permanent chemical modifier in simultaneous atomic absorption spectrometry: interference studies on As, Cd, Pb and Se determination
- Source :
- Journal of the Brazilian Chemical Society, Vol 15, Iss 1, Pp 75-82 (2004)
- Publication Year :
- 2004
- Publisher :
- Sociedade Brasileira de QuĂmica, 2004.
-
Abstract
- A systematic study of Na+, K+, Ca2+, Mg2+, Cl-, NO3-, SO4(2-) and PO4(3-) interference on the simultaneous atomization of As, Cd, Pb and Se was carried out using an electrothermal atomic absorption spectrometer. The mixture 250 mg W + 250 mg Rh was thermally deposited on the graphite tube platform and used as permanent chemical modifier. The compromised heating program adjustments (Tp = 650&ordmC and Ta = 2200&ordmC) diminished the thermal decomposition and volatilization efficiency of the interfering species during the pyrolysis step. The Cl- and SO4(2-) anions caused the most severe interference, while NO3- only affected the analytes' atomization when high amounts (> 27 mg) were used. The anion PO4(3-) produced a strong negative interference on the Se absorbance signals. The presence of Na+, K+, Ca2+ and Mg2+ prejudiced mainly Cd, and the interference effects of Cl- salts were more pronounced than NO3- salts.