- Objective: Pulse Wave Velocity (PWV) can be measured between different sites. Here we used two different aspects to assess the PWV; the standard method Carotid-Femoral (CF) Sphygmocor (AtCorMedical – Australia) and the pOpmetre ® (Axelife SAS – France) which uses the Finger to Toe (FT) signals. The aim of this study was to evaluate the agreement between FT-PWV and CF-PWV and to test the robustness of pOpmetre ® build-in algorithm. Design and method: CF and FT PWV was measured in 150 subjects. MatLab was used to calculate FT-PWV from pOpmetre ® waveforms using four methods: maximum of second derivative (used by pOpmetre ®), intersecting tangents, 10% threshold and the cross correlation method. Results: Using built in algorithms, the comparisons of the PWVs and transit times showed a good agreement between the two methods. FT-PWV correlated with CF-PWV (r2=0.51; p