Characterising and tackling thermally induced zero-drift in displacement measuring interferometry using temperature-controlled enclosure

Authors :
Rerucha, Simon
Hola, Miroslava
Cip, Ondrej
Lazar, Josef
Oulehla, Jindrich
Mikel, Bretislav
Source :
Meas. Sci. Technol. 35 115011 (2024)
Publication Year :
2024

Abstract

Our research efforts in displacement measurement interferometry focused on long-term drifts initiated an extended experimental investigation in the interferometric assemblies of our design. We aimed to analyze, characterize and tackle the long-term measurement stability, expressed as the zero-drift, with particular attention to the thermal effects. For the experimentation, we developed a thermostatic chamber equipped with active temperature regulation, an array of sensors and control electronics. With either the finely stabilized temperature or with the thermal cycling, we can carry out a range of investigations: verification of modified design or prototype interferometers, testing of production pieces, characterization of integrated assemblies and units in terms of the zero drift and the susceptibility to thermal effects -- the temperature sensitivity ($\delta L / \delta T$, expressed in nm.K$^{-1}$). With these experimental studies, we demonstrate the potential of the zero drift studies to contribute to the development and broader expansion of interferometric instrumentation.<br />Comment: 23 pages, 2 tables, 9 figures. Published in Measurement Science and Technology. Note: This is the version of the article before peer review or editing, as submitted by an author to Measurement Science and Technology. IOP Publishing Ltd is not responsible for any errors or omissions in this version of the manuscript or any version derived from it

Details

Database :
arXiv
Journal :
Meas. Sci. Technol. 35 115011 (2024)
Publication Type :
Report
Accession number :
edsarx.2404.06184
Document Type :
Working Paper
Full Text :
https://doi.org/10.1088/1361-6501/ad646b