Back to Search Start Over

First (calibration) experiment using proton beam from FRENA at SINP

Authors :
Basu, C.
Banerjee, K.
Ghosh, T. K.
Mukherjee, G.
Bhattacharya, C.
Desai, Shraddha S
Shil, R.
Saha, A. K.
Meena, J. K.
Bar, T.
Basak, D.
Sahoo, L. K.
Saha, S.
Marick, C.
Das, D.
Kujur, M.
Roy, S.
Basu, S. S.
Gond, U.
Saha, A.
Das, A.
Samanta, M.
Saha, P.
Karan, S. K.
Publication Year :
2024

Abstract

This work presents the first calibration experiment of a 3 MV Tandetron accelerator, FRENA, performed in May 2022. The $^7$Li(p,n) reaction threshold was measured to calibrate the terminal voltage measuring device. A LiF target of thickness 175 $\mu$g/cm$^2$ was used in the experiment. The measured threshold was 1872$\pm$2.7 keV, indicating 6$-$10 keV energy shift.