Gain stabilization and consistency correction approach for multiple SiPM-based gamma-ray detectors on GECAM

Authors :
Zhang, Dali
Li, Xinqiao
Wen, Xiangyang
Xiong, Shaolin
An, Zhenghua
Sun, Xilei
Qiao, Rui
Li, Zhengwei
Gong, Ke
Hou, Dongjie
Li, Yanguo
Liang, Xiaohua
Liu, Xiaojing
Liu, Yaqing
Peng, Wenxi
Yang, Sheng
Zhang, Fan
Zhao, Xiaoyun
Cai, Ce
Li, Chaoyang
Liu, Jiacong
Xiao, Shuo
Wang, Chenwei
Yi, Qibin
Zheng, Chao
Publication Year :
2021

Abstract

Each satellite of the Gravitational wave high-energy Electromagnetic Counterpart All-sky Monitor (GECAM, mission) consists of 25 SiPM based gamma-ray detectors (GRDs). Although SiPM based GRD has merits of compact size and low bias-voltage, the drift of the SiPM gain with temperature is a severe problem for GRD performance. An adaptive voltage supply source was designed to automatically adjust the SiPM bias voltage to compensate the temperature effects and keep the gain stable. This approach has been proved to be effective during both the on-ground and in-flight tests. The in-flight measured variation of the SiPM gain is within 2%. To reduce the gain non-uniformity of GRDs, an iterative bias voltage adjustment approach is proposed and implemented. The gain non-uniformity is reduced from 17% to 0.6%. In this paper, the gain stabilization and consistency correction approach are presented and discussed in detail.<br />Comment: 14 pages, 21 figures