Test of a prototype neutron spectrometer based on diamond detectors in a fast reactor

Authors :
Osipenko, M.
Pompili, F.
Ripani, M.
Pillon, M.
Ricco, G.
Caiffi, B.
Cardarelli, R.
Verona-Rinati, G.
Argiro, S.
Publication Year :
2015

Abstract

A prototype of neutron spectrometer based on diamond detectors has been developed. This prototype consists of a $^6$Li neutron converter sandwiched between two CVD diamond crystals. The radiation hardness of the diamond crystals makes it suitable for applications in low power research reactors, while a low sensitivity to gamma rays and low leakage current of the detector permit to reach good energy resolution. A fast coincidence between two crystals is used to reject background. The detector was read out using two different electronic chains connected to it by a few meters of cable. The first chain was based on conventional charge-sensitive amplifiers, the other used a custom fast charge amplifier developed for this purpose. The prototype has been tested at various neutron sources and showed its practicability. In particular, the detector was calibrated in a TRIGA thermal reactor (LENA laboratory, University of Pavia) with neutron fluxes of $10^8$ n/cm$^2$s and at the 3 MeV D-D monochromatic neutron source named FNG (ENEA, Rome) with neutron fluxes of $10^6$ n/cm$^2$s. The neutron spectrum measurement was performed at the TAPIRO fast research reactor (ENEA, Casaccia) with fluxes of 10$^9$ n/cm$^2$s. The obtained spectra were compared to Monte Carlo simulations, modeling detector response with MCNP and Geant4.<br />Comment: ANIMMA 2015 Conference proceedings, 20-24 April 2015, Lisbon, Portugal

Details

Database :
arXiv
Publication Type :
Report
Accession number :
edsarx.1505.06314
Document Type :
Working Paper