Back to Search Start Over

Characterization of a commercial, front-illuminated interline transfer CCD camera for use as a guide camera on a balloon-borne telescope

Authors :
Clark, Paul
Massey, Richard
Chang, Herrick L.
Galloway, Mathew
Israel, Holger
Jones, Laura L.
Li, Lun
Mandic, Milan
Morris, Tim
Netterfield, Barth
Peacock, John
Sharples, Ray
Susca, Sara
Publication Year :
2014

Abstract

We report results obtained during the characterization of a commercial front-illuminated progressive scan interline transfer CCD camera. We demonstrate that the unmodified camera operates successfully in temperature and pressure conditions (-40C, 4mBar) representative of a high altitude balloon mission. We further demonstrate that the centroid of a well-sampled star can be determined to better than 2% of a pixel, even though the CCD is equipped with a microlens array. This device has been selected for use in a closed-loop star-guiding and tip-tilt correction system in the BIT-STABLE balloon mission.<br />Comment: 14 pages, 16 figures. Accepted in Journal of Astronomical Instrumentation, special issue on Scientific Ballooning

Details

Database :
arXiv
Publication Type :
Report
Accession number :
edsarx.1408.6529
Document Type :
Working Paper