Back to Search Start Over

Sources of the systematic errors in measurements of Po-214 decay half-life time variations at the Baksan deep underground experiments

Authors :
Alexeyev, E. N.
Gavrilyuk, Yu. M.
Gangapshev, A. M.
Kazalov, V. V.
Kuzminov, V. V.
Panasenko, S. I.
Ratkevich, S. S.
Publication Year :
2014

Abstract

The design changes of the Baksan low-background TAU-1 and TAU-2 set-ups allowed to improve a sensitivity of Po-214 half-life (\tau) measurements up to the 2.5 x 10^{-4} are described. Different possible sources of systematic errors influencing on the $\tau$-value are studed. An annual variation of Po-214 half-life time measurements with an amplitude of A=(6.9 \pm 3) x 10^{-4} and a phase of (\phi=93 \pm 10) days was found in a sequence of the week-collected \tau-values obtained from the TAU-2 data sample with total duration of 480 days. 24 hours' variation of the \tau-value measurements with an amplitude of A=(10.0 \pm 2.6) x 10^{-4} and phase of (\phi=1 \pm 0.5) hours was found in a solar day 1 hour step \tau-value sequence formed from the same data sample. It was found that the Po-214 half-life averaged at 480 days is equal to (163.45 \pm 0.04) mks.<br />Comment: 17 pages, 6 figures, accepted for publication in Journal of Physics of Elementary Particles and Atomic Nuclei

Details

Database :
arXiv
Publication Type :
Report
Accession number :
edsarx.1404.5769
Document Type :
Working Paper
Full Text :
https://doi.org/10.1134/S1063779615020021