Back to Search Start Over

X-ray radiographic technique for measuring density uniformity of silica aerogel

Authors :
Tabata, Makoto
Hatakeyama, Yoshikiyo
Adachi, Ichiro
Morita, Takeshi
Nishikawa, Keiko
Source :
Nucl. Instr. and Meth. A 697 (2013) 52
Publication Year :
2012

Abstract

This paper proposes a new X-ray radiographic technique for measuring density uniformity of silica aerogels used as radiator in proximity-focusing ring-imaging Cherenkov detectors. To obtain high performance in a large-area detector, a key characteristic of radiator is the density (i.e. refractive index) uniformity of an individual aerogel monolith. At a refractive index of n = 1.05, our requirement for the refractive index uniformity in the transverse plane direction of an aerogel tile is |\delta (n - 1)/(n - 1)| < 4% in a focusing dual layer radiator (with different refractive indices) scheme. We applied the radiographic technique to evaluate the density uniformity of our original aerogels from a trial production and that of Panasonic products (SP-50) as a reference, and to confirm they have sufficient density uniformity within \pm 1% along the transverse plane direction. The measurement results show that the proposed technique can quantitatively estimate the density uniformity of aerogels.<br />Comment: To be published in Nucl. Instr. and Meth. A, 8 pages, 6 figures, 1 table

Details

Database :
arXiv
Journal :
Nucl. Instr. and Meth. A 697 (2013) 52
Publication Type :
Report
Accession number :
edsarx.1212.3560
Document Type :
Working Paper
Full Text :
https://doi.org/10.1016/j.nima.2012.09.001