Back to Search
Start Over
Magnetocardiography with a modular spin-exchange relaxation free atomic magnetometer array
- Source :
- 2012 Phys. Med. Biol. 57 2619
- Publication Year :
- 2011
-
Abstract
- We present a portable four-channel atomic magnetometer array operating in the spin exchange relaxation-free regime. The magnetometer array has several design features intended to maximize its suitability for biomagnetic measurement, specifically foetal magnetocardiography, such as a compact modular design, and fibre coupled lasers. The modular design allows the independent positioning and orientation of each magnetometer, in principle allowing for non-planar array geometries. Using this array in a magnetically shielded room, we acquire adult magnetocadiograms. These measurements were taken with a 6-11 fT Hz^(-1/2) single-channel baseline sensitivity that is consistent with the independently measured noise level of the magnetically shielded room.<br />Comment: 15 pages, 5 figures
Details
- Database :
- arXiv
- Journal :
- 2012 Phys. Med. Biol. 57 2619
- Publication Type :
- Report
- Accession number :
- edsarx.1106.4779
- Document Type :
- Working Paper
- Full Text :
- https://doi.org/10.1088/0031-9155/57/9/2619