Back to Search
Start Over
Characterization of an INVS Model IV Neutron Counter for High Precision ($\gamma,n$) Cross-Section Measurements
- Source :
- Nucl.Instrum.Meth.A647:55-62,2011
- Publication Year :
- 2010
-
Abstract
- A neutron counter designed for assay of radioactive materials has been adapted for beam experiments at TUNL. The cylindrical geometry and 60% maximum efficiency make it well suited for ($\gamma,n$) cross-section measurements near the neutron emission threshold. A high precision characterization of the counter has been made using neutrons from several sources. Using a combination of measurements and simulations, the absolute detection efficiency of the neutron counter was determined to an accuracy of $\pm$ 3% in the neutron energy range between 0.1 and 1 MeV. It is shown that this efficiency characterization is generally valid for a wide range of targets.<br />Comment: 22 pages, 13 figures
- Subjects :
- Nuclear Experiment
Subjects
Details
- Database :
- arXiv
- Journal :
- Nucl.Instrum.Meth.A647:55-62,2011
- Publication Type :
- Report
- Accession number :
- edsarx.1010.3930
- Document Type :
- Working Paper
- Full Text :
- https://doi.org/10.1016/j.nima.2011.01.177