Back to Search Start Over

Simulations of a Scintillator Compton Gamma Imager for Safety and Security

Authors :
Sinclair, L. E.
Hanna, D. S.
MacLeod, A. M. L.
Saull, P. R. B.
Source :
IEEE Trans.Nucl.Sci.56:1262-1268,2009
Publication Year :
2008

Abstract

We are designing an all-scintillator Compton gamma imager for use in security investigations and remediation actions involving radioactive threat material. To satisfy requirements for a rugged and portable instrument, we have chosen solid scintillator for the active volumes of both the scatter and absorber detectors. Using the BEAMnrc/EGSnrc Monte Carlo simulation package, we have constructed models using four different materials for the scatter detector: LaBr_3, NaI, CaF_2 and PVT. We have compared the detector performances using angular resolution, efficiency, and image resolution. We find that while PVT provides worse performance than that of the detectors based entirely on inorganic scintillators, all of the materials investigated for the scatter detector have the potential to provide performance adequate for our purposes.<br />Comment: Revised text and figures, Presented at SORMA West 2008, Published in IEEE Transactions on Nuclear Science

Details

Database :
arXiv
Journal :
IEEE Trans.Nucl.Sci.56:1262-1268,2009
Publication Type :
Report
Accession number :
edsarx.0809.4240
Document Type :
Working Paper
Full Text :
https://doi.org/10.1109/TNS.2009.2019271