Back to Search Start Over

Evaluation of a research use only luminex based assay for measurement of procalcitonin in serum samples

Authors :
Garrigan, C.
Han, J.
Tolomeo, P.
Johnson, K. J.
Stephen Master
Lautenbach, E.
Nachamkin, I.
Source :
Scopus-Elsevier
Publication Year :
2016
Publisher :
e-Century Publishing Corporation, 2016.

Abstract

Research use only (RUO) assays do not undergo a validation process similar to test kits used for clinical purposes. Several studies have suggested that RUO assays need to be validated prior to use in any research studies. We evaluated a research use only Luminex platform based assay for measuring serum procalcitonin levels (Bio-Plex ProTM Human Acute Phase Multiplex Assay, Bio-Rad Laboratories, Hercules, CA) for comparability with an FDA cleared assay for procalcitonin (VIDAS B.R.A.H.M.S. PCT Assay, bioMérieux, Durham, NC). We tested 1,072 serum samples collected from patients with suspected sepsis in an intensive care unit setting for the comparison. There was poor correlation of the luminex based assay (r=0.081) with the VIDAS PCT Assay in the clinically relevant measurement range (

Details

Language :
English
Database :
OpenAIRE
Journal :
Scopus-Elsevier
Accession number :
edsair.pmid.dedup....7d3a830b78a57e9d4ef2f6cfb09ca8af