Multicenter evaluation of the Paragon CZE 2000 capillary zone electrophoresis system for serum protein electrophoresis and monoclonal component typing

Authors :
J, Bienvenu
M S, Graziani
F, Arpin
H, Bernon
C, Blessum
C, Marchetti
G, Righetti
M, Somenzini
G, Verga
F, Aguzzi
Source :
Clinical chemistry. 44(3)
Publication Year :
1998

Abstract

Serum protein electrophoresis and typing of monoclonal components (MCs) are routine but time-consuming and technically demanding assays. We evaluated capillary electrophoresis (Paragon CZE 2000) for automation of the two assays. CZE and cellulose acetate electrophoresis gave similar data on 794 samples. Within-run and between-run CVs were2% for albumin and gamma-globulins and 4-7% for alpha 1-, alpha 2-, and beta-globulins. Bilirubin, hemoglobin, triglycerides, and fibrinogen were found not to interfere. No carryover by capillaries was detected. The detection limit for MC was0.5 g/L. MC assessment by immunosubtraction on 403 samples identified the monoclonal type in all samples with peak concentrations10 g/L; only 50% of MCs that could not be quantified by densitometric scan were typed.