Retaining of PTCA guide wire in the left ventricular lead and subsequent application of epicardial electrode when CRT-D implantation in a patient with severe heart failure and persistent left superior vena cava: a case report

Authors :
Hou, Pei-Pei
Liu, Yu-Hong
Qu, Hai-Bo
Meng, Jin
Li, Qiang
Miao, Zhi-Lin
Publication Year :
2015
Publisher :
e-Century Publishing Corporation, 2015.

Abstract

Objective: One patient with severe heart failure (LV 92 mm, EF 28%) was treated by cardiac resynchronization therapy (CRT). Method: During the operation, it was found that double superior vena cava coexisted, and selective coronary venography cannot clearly show every branch. It was difficult to push ventriculus sinister electrode to sideward vein, so the electrode was released to far end of frontal septal branch along great cardiac vein. Result: However, because of insufficient braced force of ventriculus sinister electrode, 0.014 PTCA guide wire was detained in the electrode. 2 years later, two spots of PTCA guide wire retained in ventriculus sinister electrode broke in atrium dextrum, so the implantation of epicardial electrode was conducted. Conclusion: After the operation, heart failure was relieved. After 43 months, the battery of pacemaker depleted, so the pacemaker was changed. The effect since follow-up visit was good, LV decreased to 86 mm, EF increased to 32%, and SPWMD time limit shortened from 147 ms to 45 ms. The therapeutic experience of this patient indicated that the effect of detaining PTCA guide wire to enhance braced force in implantation of ventriculus sinister is unreliable and inappropriate to be advocated.

Details

Language :
English
Database :
OpenAIRE
Accession number :
edsair.pmid..........e85debdfc2a4b3da370618aa1fcfb0c7