Performance evaluation of the Abbott Cell-Dyn 1800 automated hematology analyzer

Authors :
Richard, Kendall
Ellen, Benoit
Jean, Bogiages
Richard, Bordenkircher
Kim, Caple
Ling-Ling, Chen
Tony, Cheng
Tina, Hoshino
Jeffrey, Kelley
Ngoc, Ngo
Thomas, Schisano
Philip, Stevenson
Caroline, Tsou
J P, Yang
Source :
Laboratory hematology : official publication of the International Society for Laboratory Hematology. 9(3)
Publication Year :
2003

Abstract

The Cell-Dyn 1800 is a new automated hematology analyzer that provides an 18-parameter blood count and operates at a throughput of 60 specimens per hour. (Clinical significance has not been established for plateletcrit and platelet distribution width; therefore these parameters are not reportable in the United States.) The instrument is smaller than its predecessor and offers a variety of enhanced features, including increased patient data storage, availability of a bar code reader, and a cyanide-free reagent system. Our study comprehensively assessed the analytical characteristics of the instrument in a variety of areas. When applicable the requirements of appropriate NCCLS guideline documents were followed. Instrument background and carryover were shown to be consistent with the performance specification. Accuracy was demonstrated by linearity studies and by comparison of results from the automated blood count parameters of the Cell-Dyn 1700 and from the manual differential method. The instrument-generated results and morphology flags were found to be effective in screening for abnormalities demonstrated by manual morphological assessment. Precision of the Cell-Dyn 1800 instrument was studied over the short term (with fresh blood) as well as the long term with stabilized quality control material. In either case, the instrument exhibited satisfactory performance. The blood count parameters showed acceptable stability during storage experiments, although in the case of mean platelet volume, the Cell-Dyn 1800 results increased during the first 6 hours of storage, reflecting the morphological changes expected of platelets from EDTA-anticoagulated specimens.

Details

ISSN :
10802924
Volume :
9
Issue :
3
Database :
OpenAIRE
Journal :
Laboratory hematology : official publication of the International Society for Laboratory Hematology
Accession number :
edsair.pmid..........e23befbce04876eacdbd2d4a2fd15a1a