In vitro evaluation of the hemostatic effect of method involving the combined use of Hydrofit
- Source :
- General thoracic and cardiovascular surgery. 68(9)
- Publication Year :
- 2019
-
Abstract
- We developed an effective hemostatic method using HydrofitA simulated circuit was created using the pump of a Nipro ventricular assist system and a prosthetic graft. A hole was made in the graft by a needle and three hemostatic methods were applied: the silicone sheet method (SS) using HydrofitIn the 60 s compression, the amount of leakage after SS, BB, and FT was 0.4 ± 0.8, 0.2 ± 0.6, and 0 ± 0.0 ml, respectively, and FT showed no leakage. In the 30 s compression, the amount of leakage after SS, BB, and FT was 14.2 ± 27.9, 1.0 ± 3.2, and 7.8 ± 22.6 ml, respectively, and did not differ to a statistically significant extent.The method of combining Hydrofit