Back to Search
Start Over
Intraoperative Evaluation of Laparoscopic Insufflation Technique for Quality Control in the OR
- Source :
- JSLS : Journal of the Society of Laparoendoscopic Surgeons
- Publication Year :
- 2000
- Publisher :
- Society of Laparoendoscopic Surgeons, 2000.
-
Abstract
- Objective: With increasing technology and computerized systems in the OR, the physician's responsibility is growing. For intraoperative evaluation of insufflation techniques, a data acquisition model for quality control study of potential insufflation problems is necessary. Methods: A computer-based, online data acquisition model was designed with a Pentium notebook, PCMCIA data acquisition board PCI-460-P1 and a Visual Designer 3.0 measurement program (both Intelligent Instrumentation, Inc., Tucson, AZ), temperature meters Therm 2280-1 and 2283-2 (Ahlborn, Holzkirchen, Germany) and temperature probes 401 AC and 402 AC (YSI, Inc., Yellow Springs, OH) and T-430-2R (Ahlborn, Holzkirchen, Germany). Gas flow was measured with laminar flow element LFE 1 and flow meters Digima premo 720 (both Special Instruments, Noerdlingen, Germany). During 73 standard laparoscopic procedures, gas flow (L/min) in the insufflation hose, pressure (mm Hg) in the hose and abdomen as well as temperature (°C) in the hose, abdomen and rectum were measured continuously at 3 Hz rate. Results: Actual values measured show a wide range often not identical with insufflator presetting. Pressure in the abdomen is usually less than hose pressure. Intraabdominal pressure peaks (≤50 mm Hg) occurred during insufficient anesthesia, while leaning on the abdomen, during trocar insertion and other manipulation. Blood irrigation fluids found in the hose (n=3/73) can lead to bacterial contamination. Negative pressure (−50 mm Hg) was measured due to Endobag removal. Negative flow (≤15 L/min) was caused by pressure on the abdomen, insufflator regulation and an empty CO2 gas tank. Gas temperature in the hose equals room temperature but can decrease in the abdomen to 27.7°C due to high gas flow, large amounts of gas used and prolonged insufflation. Further insufflation-related problems were documented. Conclusions: This computer-based measurement model proved to be useful for quality control study in the OR. Results demonstrate the need for intraoperative evaluation of insufflation techniques for laparoscopy. Although no obvious complication related to insufflation problems occurred, some findings potentially question patient security.
Details
- Language :
- English
- ISSN :
- 19383797 and 10868089
- Volume :
- 4
- Issue :
- 3
- Database :
- OpenAIRE
- Journal :
- JSLS : Journal of the Society of Laparoendoscopic Surgeons
- Accession number :
- edsair.pmid..........d136b44d00a4cb21d80514c42d27263e