Back to Search Start Over

Use of Factory-Calibrated Real-time Continuous Glucose Monitoring Improves Time in Target and HbA

Authors :
Hood, Thabit
Joshi Navis, Prabhu
Womba, Mubita
Catherine, Fullwood
Shazli, Azmi
Andrea, Urwin
Ian, Doughty
Lalantha, Leelarathna
Source :
Diabetes care. 43(10)
Publication Year :
2020

Abstract

International type 1 diabetes registries have shown that HbAWe conducted a randomized crossover trial in young people with type 1 diabetes (16-24 years old) comparing the Dexcom G6 CGM system and self-monitoring of blood glucose (SMBG). Participants were assigned to the interventions in random order during two 8-week study periods. During SMBG, blinded continuous glucose monitoring (CGM) was worn by each participant for 10 days at the start, week 4, and week 7 of the control period. HbATime in range was significantly higher during CGM compared with control (35.7 ± 13.5% vs. 24.6 ± 9.3%; mean difference 11.1% [95% CI 7.0-15.2];CGM use in young people with type 1 diabetes improves time in target and HbA