Back to Search Start Over

[Performance evaluation of Mindray BS-200 automatic biochemistry analyzer]

Authors :
Bing, Xie
Hany-uan, Zhang
Li-pin, Jiang
Jun, Wang
Chen-ying, Dai
Ya-qing, Pang
Source :
Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation. 33(3)
Publication Year :
2009

Abstract

To evaluate the main performance of BS-200 automatic biochemistry analyzer.To evaluate the repeatability, linearity, precision and accurate Of the BS-200 automatic biochemistry analyzer by 8 items that contain all wavelength (ALT, GGT, Cl-, Cr, Glu, TP, Alb. APOA1. TBil), and carry on the contrast experiment with Hitachi 7170.The daytime run CV,within-run CV between-run CV and total CV are under 4.0%, the duplication is well, the accuracy is high, the linear scope is wide, and the contrast experiment correlation is good.This instrument each appraisal target is good, and worth in the small or middle-sized hospital promotion application.

Details

ISSN :
16717104
Volume :
33
Issue :
3
Database :
OpenAIRE
Journal :
Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
Accession number :