Reducing arrhythmias associated with central venous catheter insertion or exchange
- Source :
- Nutrition (Burbank, Los Angeles County, Calif.). 8(1)
- Publication Year :
- 1992
-
Abstract
- A recent study demonstrated that the incidence of new arrhythmias occurring during central venous catheter insertion or exchange was 41% atrial and 25% ventricular arrhythmias (12% couplets or greater). Over-insertion of the guidewire, causing direct stimulation to the right side of the heart, has been postulated to be the causative factor. A new technique that allows the operator to control the length of guidewire inserted was developed. With this technique on a population of hospitalized patients, similar to those in the previous study, the incidence of atrial arrhythmias decreased to 32% and the incidence of ventricular arrhythmias to 6% (single premature ventricular contractions only). Although this new technique has limitations, there was a dramatic improvement in the incidence of cardiac arrhythmias. These results indicate a need for modifications in the available equipment to avoid the infrequent but life-threatening complication of malignant arrhythmia.