Back to Search
Start Over
Immunofluorescence microtip sensor for point-of-care tuberculosis (TB) diagnosis
- Source :
- Methods in molecular biology (Clifton, N.J.). 1256
- Publication Year :
- 2015
-
Abstract
- A immunofluorescence microtip sensor was developed for specific detection of Mycobacterium cells in sputum samples by the combination of electric field, streaming flow, and immuno-affinity binding. The detection limit was 200 CFU/mL in human sputum, which was comparable to PCR but without requiring bacteriological culture, centrifugation, or nucleic acid amplification. In spite of the complex nature of physical, chemical, and biological mechanisms, the simple operation of "dipping and withdrawal" of tips will allow for screening by minimally trained personnel within 30 min. In addition, the minimal power requirement (5 W) combined with low assay cost is ideal for point-of-care (POC) screening in resource-limited settings.
Details
- ISSN :
- 19406029
- Volume :
- 1256
- Database :
- OpenAIRE
- Journal :
- Methods in molecular biology (Clifton, N.J.)
- Accession number :