Back to Search Start Over

[Evaluation of the FREEDOM O2, DC Concentrator, a portable oxygen concentrator capable of operating under car battery power]

Authors :
T, Nakamoto
T, Nakamachi
T, Hase
S, Okamoto
I, Kisiro
H, Kuga
N, Onuma
H, Harasawa
S, Kato
N, Kaneko
Source :
Nihon Kokyuki Gakkai zasshi = the journal of the Japanese Respiratory Society. 36(7)
Publication Year :
1998

Abstract

The FREEDOM O2 DC Concentrator, an oxygen concentrator which can be powered by a car battery, was evaluated. The oxygen concentrator was used by a 67 year-old man with sequelae of pulmonary tuberculosis who was receiving long-term oxygen therapy at home (HOT), and whose work required automobile trips over long distances. The equipment used was an adsorption-type oxygen concentrator capable of operating on a DC 12 V power supply, and which can be powered from a residential power outlet (AC 100 V) using a dedicated voltage converter. The trunk-shaped equipment measured 21/584 x 42 cm and weighed 17 kg. The noise level of the equipment was 58.2 +/- 2.5 dB (at 1 meter), and the flow rate can be set to 0.25, 0.50, 0.75, 1.0, 1.5, and 2.0 l/min.1) The O2 concentration which can be generated by this equipment is 93 +/- 3% (0.25 to 1.5 l/min) or 90 +/- 2.8% (2.0 l/min). 2) Using this equipment, the patient was capable of driving himself in comfort for two hours or longer. Further, it was possible to stay in a hotel during a trip, inhaling oxygen generated by the equipment. Hereafter, this equipment should enable or facilitate long-distance driving, travel and lodging for HOT patients.

Details

ISSN :
13433490
Volume :
36
Issue :
7
Database :
OpenAIRE
Journal :
Nihon Kokyuki Gakkai zasshi = the journal of the Japanese Respiratory Society
Accession number :
edsair.pmid..........7274fd35e7c764ca0b6109562e5cf395