MAGORINO: Magnitude-only fat fraction and R

Authors :
Timothy J P, Bray
Alan, Bainbridge
Emma, Lim
Margaret A, Hall-Craggs
Hui, Zhang
Source :
Magnetic resonance in medicineREFERENCES. 89(3)
Publication Year :
2022

Abstract

Magnitude-based fitting of chemical shift-encoded data enables proton density fat fraction (PDFF) andSimulations of multi-echo gradient-echo signal intensities are used to investigate the performance and behavior of MAGORINO over the space of clinically plausible PDFF,Simulations show that Rician noise-based magnitude fitting outperforms existing Gaussian noise-based fitting and reveals two key mechanisms underpinning the observed improvement. First, the likelihood functions exhibit two local optima; Rician noise modeling increases the chance that the global optimum corresponds to the ground truth. Second, when the global optimum corresponds to ground truth for both noise models, the optimum from Rician noise modeling is closer to ground truth. Multisite phantom experiments show good agreement of MAGORINO PDFF with reference values, and in vivo experiments replicate the performance benefits observed in simulation.The MAGORINO algorithm reduces Rician noise-related bias in PDFF and