Deformation T-Cup: a new multi-anvil apparatus for controlled strain-rate deformation experiments at pressures above 18 GPa

Authors :
Simon A, Hunt
Donald J, Weidner
Richard J, McCormack
Matthew L, Whitaker
Edward, Bailey
Li, Li
Michael T, Vaughan
David P, Dobson
Source :
The Review of scientific instruments. 85(8)
Publication Year :
2014

Abstract

A new multi-anvil deformation apparatus, based on the widely used 6-8 split-cylinder, geometry, has been developed which is capable of deformation experiments at pressures in excess of 18 GPa at room temperature. In 6-8 (Kawai-type) devices eight cubic anvils are used to compress the sample assembly. In our new apparatus two of the eight cubes which sit along the split-cylinder axis have been replaced by hexagonal cross section anvils. Combining these anvils hexagonal-anvils with secondary differential actuators incorporated into the load frame, for the first time, enables the 6-8 multi-anvil apparatus to be used for controlled strain-rate deformation experiments to high strains. Testing of the design, both with and without synchrotron-X-rays, has demonstrated the Deformation T-Cup (DT-Cup) is capable of deforming 1-2 mm long samples to over 55% strain at high temperatures and pressures. To date the apparatus has been calibrated to, and deformed at, 18.8 GPa and deformation experiments performed in conjunction with synchrotron X-rays at confining pressures up to 10 GPa at 800 °C .