Back to Search
Start Over
Mechanical failure of the Björk-Shiley valve. Updated follow-up and considerations on prophylactic rereplacement
- Source :
- The Journal of thoracic and cardiovascular surgery. 97(1)
- Publication Year :
- 1989
-
Abstract
- The early production series of the 29 to 31 mm 70-degree convexo-concave (CC70) Björk-Shiley heart valves constitutes a subgroup of CC70 values with an extra high risk of outlet strut fracture. The hazard function for mechanical failure among these valves has remained almost constant over a 7-year follow-up period, and the 7-year actuarial incidence of mechanical failure among these valves is 12.5%. Selected patients with these valves will be considered for prophylactic rereplacement.
- Subjects :
- Aortic Valve
Heart Valve Prosthesis
Humans
Mitral Valve
Prosthesis Failure
Subjects
Details
- ISSN :
- 00225223
- Volume :
- 97
- Issue :
- 1
- Database :
- OpenAIRE
- Journal :
- The Journal of thoracic and cardiovascular surgery
- Accession number :
- edsair.pmid..........18cc77f9d4dec470496cd6a9f0349dc6