Estimulação ventricular bifocal no tratamento da insuficiência cardíaca com miocardiopatia dilatada
- Source :
- Brazilian Journal of Cardiovascular Surgery, Volume: 15, Issue: 1, Pages: 44-54, Published: MAR 2000
- Publication Year :
- 2000
- Publisher :
- Sociedade Brasileira de Cirurgia Cardiovascular, 2000.
-
Abstract
- INTRODUÇÃO: O alargamento do QRS pela estimulação ventricular artificial (EVa) convencional prejudica a contratilidade cardíaca e a função mitral. OBJETIVO: Propôr um modo alternativo de EVa com QRS mais estreito, com 2 eletrodos em ventrículo direito(VD) e testá-lo frente à EVa convencional. CASUÍSTICA E MÉTODOS: Cinco pacientes com indicação para implantação de marcapasso definitivo(MP), 4M/1F, idade média de 52,2 anos, portadores de ICC (III/IV) por miocardiopatia dilatada (Chagas 4, Indeterminada 1), AC 3+/4+, FA crônica com BAV, foram submetidos a implante de MP endocárdico com 2 eletrodos em VD, o primeiro na área do feixe de His (septal) e o segundo, convencional, na ponta do VD. O gerador foi Biotronik Dromos DR (único com AV mínimo de 15ms). O eletrodo septal foi conectado à saída atrial e o de ponta à saída ventricular. Após 2 semanas, os pacientes foram avaliados ecocardiograficamente, na mesma sessão, em 3 modos de estimulação na mesma freqüência: modo "AAI" (septal), modo VVI (convencional) e modo "DDT" com intervalo AV=15ms (estimulação bifocal no VD quase simultânea). RESULTADOS: As médias dos parâmetros estudados estão na tabela abaixo: CONCLUSÃO: Sob o ponto de vista ecocardiográfico, a comparação imediata no mesmo paciente dos 3 modos de estimulação revelou: 1-O modo convencional foi o de pior rendimento hemodinâmico; 2-O modo bifocal mostrou o melhor resultado com aumento médio da FE em 6,8% e do DC em 0,6 l/min, redução média do AE em 7,5 cm2, redução média da área de refluxo mitral em 7,6 cm2 e redução média na duração do QRS em 61 ms, (p < 0,03), sendo a melhor EVa na miocardiopatia dilatada com insuficiência cardíaca, insuficiência mitral funcional e indicação de MP. BACKGROUND: Widening of QRS by ventricular conventional pacing (CP) impairs ventricular contractility and mitral performance. PURPOSE: To study an alternative mode of cardiac stimulation with narrower QRS, using 2 leads in the right ventricle and to test it, compared to CP. MATERIAL AND METHODS: 5 dilated cardiomyopathy patients (3 Chagas´ disease, 2 unknown) with cardiac failure (NYHA III/4) and pacemaker (PM) indication, 4 male/1 female, mean age 52.2. years, cardiac shape 3+/4+, chronic AF with AVB, were submitted to endocardial PM implants with 2 right ventricular leads. The first in the His bundle area (septal) and the second, conventional, in the right ventricular apex. The generator was Biotronik Dromos DR (the short programmable AV = 15ms). The septal lead was connected to the atrial outflow and the one of the apex to the ventricular exit. After 2 weeks, they were studied by echocardiogram, in 3 stimulation modes in the same session with same cardiac rate: mode "AAI" (septal), mode VVI (conventional) and mode "DDT" with AV interval=15ms (almost simultaneous bifocal stimulation). RESULTS: The averages of the parameters studied were: CONCLUSION: The echocardiographic evaluation for the immediate comparison in the same patient with 3 stimulation modes, revealed: 1. The conventional mode showed the worse hemodynamic performance; 2. the bifocal mode showed the best result with mean increase of the EF of 6.8% and of CO in 0.6 l/min, mean reduction of LA in 7.5 cm2, mean reduction of the area of mitral regurgitation in 7.6 cm2 and mean reduction in the QRS duration of 61ms, (p < 0.03). It was the best stimulation in cardiomyopathy with heart failure, functional mitral regurgitation and pacemaker indication.
Details
- Language :
- Portuguese
- Database :
- OpenAIRE
- Journal :
- Brazilian Journal of Cardiovascular Surgery, Volume: 15, Issue: 1, Pages: 44-54, Published: MAR 2000
- Accession number :
- edsair.od.......608..2d779ceaad09d44818d2314fdf684225