Influência das dimensões finais do stent pós-procedimento nos resultados tardios do stent EndeavorTM: análise preliminar com ultrassom intracoronário seriado

Authors :
Brito, Andre Farinelli Lima
Costa Junior, José de Ribamar
Feres, Fausto
Nakashima, Marcelo
Siqueira, Dimytri
Costa, Ricardo
Staico, Rodolfo
Abizaid, Andréa S.
Chaves, Áurea J.
Centemero, Marinella P.
Tanajura, Luiz Fernando L.
Mattos, Luiz Alberto
Maldonado, Galo
Abizaid, Alexandre
Sousa, Amanda G. M. R.
Sousa, J. Eduardo
Source :
Revista Brasileira de Cardiologia Invasiva, Volume: 17, Issue: 3, Pages: 308-313, Published: SEP 2009
Publication Year :
2009
Publisher :
Sociedade Brasileira de Hemodinâmica e Cardiologia Intervencionista - SBHCI, 2009.

Abstract

INTRODUÇÃO: Vários fatores têm sido relacionados à necessidade de revascularização da lesão-alvo, sejam eles clínicos, angiográficos ou ultrassonográficos. Quanto aos últimos, a área mínima do stent pós-procedimento (AMS) tem sido correlacionada com áreas luminais mínimas intrastent tardias (ALM), em pacientes tratados com stents farmacológicos e não-farmacológicos. OBJETIVO: Avaliar a AMS de pacientes tratados com stents liberadores de zotarolimus EndeavorTM capazes de predizer ALM < 4 mm², aos seis meses de evolução. MÉTODO: Estudo retrospectivo com 47 pacientes e 50 lesões tratados com stents liberadores de zotarolimus, submetidos a ultrassonografia intravascular (USIC) no procedimento índex e no seguimento de seis meses. Correlacionou-se a AMS com a ALM. Por meio da análise de curva ROC, avaliou-se a AMS preditora de ALM < 4 mm² no seguimento de seis meses. RESULTADOS: Observou-se correlação significativa (r² = 0,64; P = 0,001) entre AMS (6 ± 2,1 mm²) e ALM (4,6 ± 2,4 mm²). Houve diferença significativa quanto à AMS dos grupos com ALM < 4 mm² vs. > 4 mm² (4,8 mm² vs. 6,8 mm², respectivamente; P = 0,001). Pela análise da curva ROC, encontrou-se AMS pós-implante > 5,7 mm² como preditora de ALM > 4 mm² aos seis meses [área sob a curva: 0,815; intervalo de confiança (IC) 95%: 0,68-0,95; P < 0,001], com sensibilidade e especificidade de 80%. CONCLUSÃO: Em pacientes tratados com stents EndeavorTM, a AMS > 5,7 mm² esteve associada a ALM > 4 mm² no médio prazo. Tais achados podem auxiliar o implante ótimo desses stents. BACKGROUND: Several clinical, angiographic and ultrasonographic factors have been associated to stent restenosis and target lesion revascularization. As to intravascular ultrasound (IVUS) findings, the minimal stent area (MSA) has been correlated with minimal lumen areas (MLA) in patients treated with bare metal and drug eluting stents. OBJECTIVES: Evaluate MSA values of patients treated with EndeavorTM zotarolimus eluting stents capable of predicting MLA < 4 mm² after six months of follow-up. METHOD: Retrospective study with 47 patients and 50 lesions treated with zotarolimus eluting stents, submitted to IVUS at index and at the six-month follow-up. The MSA was then correlated to the MLA. ROC curve analysis was used to identify a MSA that could predict a MLA < 4 mm² at six months. RESULTS: There was a significant correlation (r² = 0.64; P = 0.001) between the MSA (6.0 ± 2.1 mm²) and MLA at six months (4.6 ± 2.4 mm²). There was a significant difference for MSA between the groups with MLA < 4 mm² vs. > 4 mm² (4.8 mm² vs. 6.8 mm², respectively; P = 0.001). ROC curve analysis identified a post-implant MSA of > 5.7 mm² as a predictor of MLA > 4 mm² at six months [area under curve: 0.815; 95% confidence interval (95% CI): 0.680.95; P < 0.001], with sensitivity and specificity of 80%. CONCLUSION: In this group of patients treated with EndeavorTM stents, MSA > 5.7 mm² was associated with MLA > 4 mm² at the mid-term follow-up. These findings may help achieve the optimal deployment of these stents.

Details

Language :
Portuguese
Database :
OpenAIRE
Journal :
Revista Brasileira de Cardiologia Invasiva, Volume: 17, Issue: 3, Pages: 308-313, Published: SEP 2009
Accession number :
edsair.od.......608..290c73adb4f7b9208a458ecd24e7eaef