Accuracy of intraocular lens calculations in eyes with keratoconus: Intraocular lens calculations in keratoconus
- Source :
- Journal of Cataract and Refractive Surgery, 49(3), 229-233. Elsevier Science
- Publication Year :
- 2023
-
Abstract
- PURPOSE: To compare the prediction accuracy of the Barrett True-K for keratoconus with standard formulas (SRK/T, Barrett Universal II and Kane) and the Kane keratoconus formula. SETTING: Shaare Zedek Medical Center, Jerusalem, Israel and University Eye Clinic, Maastricht, the Netherlands. DESIGN: Multicenter retrospective case-series. METHODS: Eyes with stable keratoconus undergoing cataract surgery were included. The predicted refractions were calculated for: SRK/T, Barrett Universal II, Barrett True-K for keratoconus (predicted and measured), Kane, and Kane adjusted for keratoconus formulas. Primary outcomes were prediction error (PE), absolute error (AE), and percentage of eyes with PE ±0.25D, ±0.50D, and ±1.00D. Subgroup analyses were performed based on the severity of the keratoconus. RESULTS: Fifty-seven eyes were included in the study. The PE was not significantly different from zero for SRK/T, Barrett True-K (predicted and measured), and Kane keratoconus formulas (range 0.09 to 0.22 D, p>0.05). The AE of Barrett True-K predicted (Median 0.14D) and Barrett True-K measured (Median 0.10D) were significantly lower from Barrett Universal II (Median 0.47D) and Kane (Median 0.50 D), p