Back to Search
Start Over
Which of 51 Plate Designs Can Most Stably Fixate the Fragments in a Fracture of the Mandibular Condyle Base?
- Source :
- Journal of Clinical Medicine; Volume 12; Issue 13; Pages: 4508
- Publication Year :
- 2023
- Publisher :
- Multidisciplinary Digital Publishing Institute, 2023.
-
Abstract
- In the surgical treatment of the most common fracture of the mandible, which is a fracture of the condylar base, a great choice of different plate shapes is observed. The aim of this study was to determine which shape gives the greatest fixation stiffness. To ensure homogeneity in comparison, tests were performed on polyurethane models divided at the level of the condylar base fracture and each were fixed with 51 plates. The plates were cut from a 1 mm thick grade 23 titanium sheet. The models were then loaded and the force required for 1 mm of fracture displacement was recorded. It was noted that in addition to osteosynthesis from two simple plates, there were also two dedicated single plates with similar rigidity. Among the large number of described designs of plates, there is considerable variation in terms of the stability of the fixation performed with them. The proposed Mechanical Excellence Factor allows a pre-evaluation of the expected rigidity of fixation with a given plate shape without the need for a loading experiment. The authors expect this to be helpful for surgeons in the application of relevant plates, as well for inventors of new plates for the osteosynthesis of basal fractures in mandibular condyle.
Details
- Language :
- English
- ISSN :
- 20770383
- Database :
- OpenAIRE
- Journal :
- Journal of Clinical Medicine; Volume 12; Issue 13; Pages: 4508
- Accession number :
- edsair.multidiscipl..7d76db92521dd54213028badae9be27b
- Full Text :
- https://doi.org/10.3390/jcm12134508