定電流型低雑音熱線風速計の試作研究

Authors :
TAKAGI, Shohei
Source :
航空宇宙技術研究所報告 = Technical Report of National Aerospace Laboratory TR-697. 697:13
Publication Year :
1982
Publisher :
航空宇宙技術研究所, 1982.

Abstract

This report represents a production of a low noise hot-wire anemometer with a constant current mode. This instrument consists of a constant current circuit, a low noise amplifier and a frequency compensator. Newly designed circuits and several techniques that reduce electric noise are described in detail. A typical application for residual turbulence measurement in a wind tunnel is given. The performance of this instrument is satisfactory for the measurement of velocity fluctuations with a turbulence intensity of less than 0.05% and with frequencies of up to 5kHz at the wind speed of 20m/s. If the heating current for the hot-wire is low, this system may be easily applied to a resistance thermometer.<br />資料番号: NALTR0697000<br />レポート番号: NAL TR-697

Details

Language :
English
ISSN :
03894010
Volume :
697
Database :
OpenAIRE
Journal :
航空宇宙技術研究所報告 = Technical Report of National Aerospace Laboratory TR-697
Accession number :
edsair.jairo.........896a0d5a6b23d3220515ce4fb9faed37