Back to Search Start Over

Determination of the main hydrolysis products of organophosphorus nerve agents, methylphosphonic acids, in human serum by indirect photometric detection ion chromatography

Authors :
Michiaki Tatsuno
Hitoshi Tsuchihashi
Mayumi Nishikawa
Munehiro Katagi
Source :
Journal of Chromatography B: Biomedical Sciences and Applications. 698:81-88
Publication Year :
1997
Publisher :
Elsevier BV, 1997.

Abstract

For the verification of the use of chemical warfare agents (CWA), sarin, soman and VX, a simple rapid and accurate method which allows us to simultaneously determine their degradation products, isopropyl methylphosphonic acid (IPMPA), pinacolyl methylphosphonic acid (PMPA), ethyl methylphosphonic acid (EMPA) and methylphosphonic acid (MPA), in human serum, was explored by indirect photometric detection ion chromatography (IPD-IC) which employs an anion-exchange column. IC analysis was performed after sample preparation with an Ag+-form cation-exchange resin cartridge, and the four methylphosphonic acids could be separated well. The proposed conditions are as follows: eluent, 0.5 mM phthalic acid-0.1 mM Tris (hydroxymethyl) aminomethane-5% acetonitrile; flow-rate, 1.0 ml/min; temperature, 50 degrees C; UV detector, 266 nm. All four methylphosphonic acids were eluted within 30 min with hardly any disturbance by impurities in the serum. Linear calibration curves were obtained for MPA, EMPA and IPMPA in the concentration range from 50 ng/ml to 1 microg/ml and for PMPA from 100 ng/ml to 1 microg/ml. The relative standard deviation for the methylphosphonic acids ranged from 3.8 to 6.9% at 500 ng/ml and the detection limits were 40 ng/ml for MPA, EMPA and IPMPA and 80 ng/ml for PMPA. The method would be suitable for analysis of human serum samples.

Details

ISSN :
03784347
Volume :
698
Database :
OpenAIRE
Journal :
Journal of Chromatography B: Biomedical Sciences and Applications
Accession number :
edsair.doi.dedup.....fe82e53a1a44aca5189cd9d1298448e2