Back to Search Start Over

Feasibility of self-gated isotropic radial late-phase MR imaging of the liver

Authors :
Robert Grimm
Petros Martirosian
Matthias Kuendel
Christer Ruff
Christina Schraml
Ahmed E. Othman
Jakob Weiss
Mike Notohamiprodjo
Jana Taron
Konstantin Nikolaou
Source :
European Radiology. 27:985-994
Publication Year :
2016
Publisher :
Springer Science and Business Media LLC, 2016.

Abstract

To evaluate feasibility of a 3D-isotropic self-gated radial volumetric interpolated breath-hold examination (VIBE) for late-phase MRI of the liver. 70 patients were included and underwent liver MRI at 1.5 T. Depending on the diagnosis, either Gd-EOB-DTPA (35 patients) or gadobutrol (35 patients) were administered. During late (gadobutrol) or hepatocyte-specific phase (Gd-EOB-DTPA), a radial prototype sequence was acquired and reconstructed using (1) self-gating with 40 % acceptance (rVIBE40); (2) with 100 % acceptance of the data (rVIBE100) and compared to Cartesian VIBE (cVIBE). Images were assessed qualitatively (image quality, lesion conspicuity, artefacts; 5-point Likert-scale: 5 = excellent; two independent readers) and quantitatively (coefficient-of-variation (CV); contrast-ratio) in axial and coronal reformations. In eight cases only rVIBE provided diagnostic image quality. Image quality of rVIBE40 was rated significantly superior (p

Details

ISSN :
14321084 and 09387994
Volume :
27
Database :
OpenAIRE
Journal :
European Radiology
Accession number :