Back to Search
Start Over
Preclinical performance of a pediatric mechanical circulatory support device: The PediaFlow ventricular assist device
- Source :
- The Journal of thoracic and cardiovascular surgery. 156(4)
- Publication Year :
- 2016
-
Abstract
- Objectives The PediaFlow (HeartWare International, Inc, Framingham, Mass) is a miniature, implantable, rotodynamic, fully magnetically levitated, continuous-flow pediatric ventricular assist device. The fourth-generation PediaFlow was evaluated in vitro and in vivo to characterize performance and biocompatibility. Methods Supported by 2 National Heart, Lung, and Blood Institute contract initiatives to address the limited options available for pediatric patients with congenital or acquired cardiac disease, the PediaFlow was developed with the intent to provide chronic cardiac support for infants as small as 3 kg. The University of Pittsburgh–led Consortium evaluated fourth-generation PediaFlow prototypes both in vitro and within a preclinical ovine model (n = 11). The latter experiments led to multiple redesigns of the inflow cannula and outflow graft, resulting in the implantable design represented in the most recent implants (n = 2). Results With more than a decade of extensive computational and experimental efforts spanning 4 device iterations, the AA battery–sized fourth-generation PediaFlow has an operating range of 0.5 to 1.5 L/min with minimal hemolysis in vitro and excellent hemocompatibility (eg, minimal hemolysis and platelet activation) in vivo. The pump and finalized accompanying implantable components demonstrated preclinical hemodynamics suitable for the intended pediatric application for up to 60 days. Conclusions Designated a Humanitarian Use Device for “mechanical circulatory support in neonates, infants, and toddlers weighing up to 20 kg as a bridge to transplant, a bridge to other therapeutic intervention such as surgery, or as a bridge to recovery” by the Food and Drug Administration, these initial results document the biocompatibility and potential of the fourth-generation PediaFlow design to provide chronic pediatric cardiac support.
- Subjects :
- Pulmonary and Respiratory Medicine
medicine.medical_specialty
medicine.medical_treatment
0206 medical engineering
Activated clotting time
Hemodynamics
02 engineering and technology
Investigational device exemption
030204 cardiovascular system & hematology
Prosthesis Design
Hemolysis
Prosthesis Implantation
03 medical and health sciences
0302 clinical medicine
Electric Power Supplies
Internal medicine
Materials Testing
medicine
Extracorporeal membrane oxygenation
Animals
Humans
Ventricular Function
Platelet activation
Sheep, Domestic
Heart Failure
Framingham Risk Score
Miniaturization
medicine.diagnostic_test
business.industry
Body Weight
Age Factors
Infant, Newborn
Infant
020601 biomedical engineering
Blood pump
Animals, Newborn
Ventricular assist device
Child, Preschool
Cardiology
Surgery
Heart-Assist Devices
Cardiology and Cardiovascular Medicine
business
Subjects
Details
- ISSN :
- 1097685X
- Volume :
- 156
- Issue :
- 4
- Database :
- OpenAIRE
- Journal :
- The Journal of thoracic and cardiovascular surgery
- Accession number :
- edsair.doi.dedup.....fb812bd0a2e95e55d4db36fdb63e53ad