Evaluation of new plate designs for symphysis pubis internal fixation

Authors :
Allan F. Tencer
John R. Schwappach
Samuel G. Agnew
Peter T. Simonian
Milton L Routt
Richard M. Harrington
Source :
The Journal of trauma. 41(3)
Publication Year :
1996

Abstract

The purpose of this study was to evaluate and compare biplanar and other newly designed plates used for pubic symphysis internal fixation to other standard plates. Our data demonstrate that neither of the newly designed symphyseal plates, curved (Zimmer four-hole plate with either two or four 4.5-mm cortical screws) nor the biplanar (Zimmer six-hole plate with four 4.5-mm cortical screws in one plane and two 4.5-mm cortical screws in another) significantly reduce motion more than the other plates tested (Synthes two-hole 4.5-mm dynamic compression plate with two 6.5-mm cancellous screws and six-hole 3.5-mm reconstruction plate with four 3.5-mm cortical screws). All of the plate constructs were able to restore motion to nearly that of the intact symphysis pubis and sacroiliac joint. In summary, using the anteroposterior compression disruption pattern and model, the disrupted symphysis does gap under load but has no effect on sacroiliac joint gapping. It does permit increase in relative flexion angles between the wings of the pelvis. All of the plate systems tested restore normal gap motion at the symphysis and normal sacroiliac joint flexion.